MNR College of Engineering & Technology

30 core concepts and 64 advanced concepts, each linked to the others it depends on.

Core concepts

The concepts this subject is built on, each with the mistake people actually make with it.

Validation against a reference standard

A device is only as good as what it was compared against, and the comparison must cover the range you intend to use.

Where people go wrong

Validating on healthy volunteers and deploying on sick patients.

Turn it into a study

Design a validation that spans the real clinical range.

Bland-Altman rather than correlation

Two methods can correlate almost perfectly and still disagree badly. Agreement needs the difference plotted against the average.

Where people go wrong

Reporting a high correlation coefficient as proof of agreement.

Turn it into a study

Re-analyse an agreement claim with the correct method.

Overfitting

A model that memorises its training data performs brilliantly there and fails on new data. Separation of training and test sets is the defence.

Where people go wrong

Reporting accuracy from the data the model learned on.

Turn it into a study

Test a published model on local data and compare performance.

Dataset shift

A model trained in one hospital meets different machines, populations and habits elsewhere, and silently degrades.

Where people go wrong

Assuming published accuracy transfers.

Turn it into a study

Measure performance of an available model on your own images.

Human factors

A device fails in use through the interface, not the circuit. Alarm fatigue and unclear controls cause real harm.

Where people go wrong

Testing the device without testing the person using it.

Turn it into a study

Observe real use and count errors and near misses.

Requirements come from the user, not the engineer

What a device must do is established by observing and asking the people who will use it, and requirements written without them describe the engineer's assumptions.

Where people go wrong

Writing requirements from a technical understanding of the problem.

Turn it into a study

Observe a clinical task and write requirements from what you saw.

Design is iterative and documented

A device design evolves through cycles of specification, build, test and revision, and the design history recording that sequence is a regulatory requirement.

Where people go wrong

Keeping design decisions in people's heads.

Turn it into a study

Set up a design history file for a project.

Change control prevents regression

A change to one part of a device can break another, which is why changes are assessed for impact, tested and documented rather than simply made.

Where people go wrong

Making a small change without impact assessment.

Turn it into a study

Define a change control process for a project.

Materials have failure modes

Fatigue, creep, corrosion and degradation each cause failure over time rather than at first use, which is why lifetime testing differs from functional testing.

Where people go wrong

Testing a device only for immediate function.

Turn it into a study

Identify the time-dependent failure modes of a material choice.

Tolerances accumulate

Each component's manufacturing tolerance adds to the next, so a device assembled from parts within specification can be out of specification.

Where people go wrong

Specifying components without a tolerance analysis.

Turn it into a study

Perform a tolerance stack-up for an assembly.

Manufacturability is a design constraint

A design that cannot be made repeatably at the required cost and volume is not a solution, which makes manufacturing method part of the design.

Where people go wrong

Designing without reference to the manufacturing process.

Turn it into a study

Review a design for manufacturability constraints.

Supply chain determines what can be built

Component availability, lead times and single-source parts constrain a design, and an obsolete component can end a product line.

Where people go wrong

Specifying a component without checking its supply.

Turn it into a study

Review a design's components for supply risk.

Validation uses the production device

A device validated as a prototype may behave differently in production, which is why validation is performed on devices made by the final process.

Where people go wrong

Validating a hand-built prototype and shipping production units.

Turn it into a study

Compare prototype and production device performance.

Labelling is regulated content

What a device's label and instructions must state is specified by regulation, including intended use, warnings and manufacturer details.

Where people go wrong

Writing labelling as marketing copy.

Turn it into a study

Review the labelling requirements for a device class.

Clinical need comes before the technology

A device that solves a problem nobody has fails whatever its elegance, which is why needs assessment with clinicians and patients precedes design.

Where people go wrong

Starting from a technology and looking for an application.

Turn it into a study

Conduct a needs assessment in a clinical area before designing.

Specifications must be measurable

A requirement that cannot be tested cannot be verified, so each specification states a quantity, a tolerance and a test method.

Where people go wrong

Writing requirements as aspirations.

Turn it into a study

Rewrite a device requirement as a testable specification.

Prototyping is for learning

Early prototypes exist to answer specific questions cheaply, and building a polished prototype before the questions are answered wastes the learning.

Where people go wrong

Building a finished-looking prototype first.

Turn it into a study

Define the question each prototype stage should answer.

Standards define the test

Device safety and performance standards specify the tests that must be passed, so reading the applicable standard early tells the designer what the device must survive.

Where people go wrong

Designing first and consulting standards for certification.

Turn it into a study

Identify the standards applicable to a device you are designing.

Measurement needs a reference

Any measuring device must be compared against a traceable reference, and accuracy claims without a stated reference and method are not claims.

Where people go wrong

Stating accuracy without the comparison method.

Turn it into a study

Define a calibration and traceability plan for a measuring device.

Signals need filtering and sampling decisions

Biosignal acquisition requires choices about sampling rate, filter bandwidth and resolution, and each choice discards information permanently.

Where people go wrong

Choosing acquisition settings by default.

Turn it into a study

Compare the same signal acquired at different settings.

Noise comes from identifiable sources

Mains interference, movement, electrode impedance and digital switching each produce characteristic noise, and identifying the source is what allows its removal.

Where people go wrong

Filtering noise without identifying its source.

Turn it into a study

Identify the noise sources in a biosignal recording.

Data pipelines need version control

Analysis that cannot be rerun on the same data to produce the same result is not reproducible, which makes version control and recorded environments part of the method.

Where people go wrong

Analysing data without recording the code and versions used.

Turn it into a study

Set up version control for an analysis and document the environment.

Testing is layered

Unit tests on components, integration tests on assemblies and system tests on the whole device catch different failures, and omitting a layer defers its failures to the patient.

Where people go wrong

Testing only the finished system.

Turn it into a study

Define tests at all three layers for a device function.

Cybersecurity applies to devices

Connected medical devices are attack surfaces, and vulnerabilities in infusion pumps and imaging systems have been demonstrated, which makes security part of safety.

Where people go wrong

Treating device security as an information technology matter.

Turn it into a study

Review the network exposure of connected devices in your hospital.

Post-market surveillance is required

A manufacturer must collect and act on field experience after approval, because problems appearing at scale were invisible in trials.

Where people go wrong

Treating approval as the end of evaluation.

Turn it into a study

Design a post-market surveillance plan for a device.

Field failures must be reported

Adverse events involving devices are reportable to regulators, and underreporting is why dangerous devices have stayed on the market for years.

Where people go wrong

Replacing a failed device without reporting.

Turn it into a study

Review device adverse event reporting in your hospital.

Documentation is part of the device

Instructions, labelling, training material and service manuals determine whether a device is used correctly, and they are regulated elements rather than accessories.

Where people go wrong

Treating documentation as an afterthought.

Turn it into a study

Review the instructions for use of a device in your hospital.

Design for the whole lifecycle

Installation, training, consumables, maintenance, repair and disposal all need designing, and a device that cannot be serviced locally fails early.

Where people go wrong

Designing for the moment of use only.

Turn it into a study

Map the full lifecycle requirements of a device design.

Intellectual property and publication interact

Publishing before filing can destroy patentability, which makes the sequence of disclosure a decision to take before writing a paper.

Where people go wrong

Presenting a device design at a conference before filing.

Turn it into a study

Review your institution's policy on disclosure and filing.

Advanced concepts

Less often taught: each one questions a common assumption, connects this subject to another, and ends in a study you could run here.

Verification versus validation

Verification asks whether you built it right; validation asks whether you built the right thing. A device can pass every technical test and be clinically useless.

Where people go wrong

Stopping at verification because it is measurable, and never testing clinical usefulness.

Turn it into a study

Design both a verification and a validation protocol for one device.

Automation bias

People accept a machine's answer even when their own judgement disagrees, and the effect is stronger when the operator is less experienced.

Where people go wrong

Deploying a decision aid without measuring whether it degrades unaided performance.

Turn it into a study

Test clinician accuracy with and without an algorithm's suggestion.

Alarm fatigue

Too many alarms, most of them false, train staff to ignore all of them, so adding an alarm can reduce safety.

Where people go wrong

Treating each alarm as a safety addition without counting the total.

Turn it into a study

Count alarms per bed per shift in your intensive care unit and the proportion actionable.

Explainability against accuracy

The most accurate models are often the least explainable, and in clinical use explanation drives trust, auditability and error detection.

Where people go wrong

Choosing on accuracy alone when the model must be defended to a patient or a court.

Turn it into a study

Compare a simple and a complex model on the same local data for both accuracy and interpretability.

Frugal and reverse innovation

Designing under severe cost and infrastructure constraints produces devices that then find markets in rich countries. The constraint is a design advantage, not only a limitation.

Where people go wrong

Treating low-resource design as a stripped-down version of an expensive device.

Turn it into a study

Take one expensive device and redesign its core function under a hundredth of the budget.

Low-resource design constraints

Intermittent power, no cold chain, dust, humidity, untrained operators and no service contract. Each is a hard design requirement, not an excuse.

Where people go wrong

Validating a device in a laboratory and assuming field performance.

Turn it into a study

Test an existing device against each constraint and record where it fails.

Signal, noise and the sampling theorem

A signal must be sampled at more than twice its highest frequency or it is irrecoverably distorted, which sets the floor for every biosignal device.

Where people go wrong

Choosing a sampling rate by convenience and producing aliased data that look plausible.

Turn it into a study

Sample one biosignal at several rates and demonstrate aliasing.

Edge computing for intermittent connectivity

Processing on the device rather than in the cloud, which is a requirement where connectivity is unreliable rather than an optimisation.

Where people go wrong

Designing for continuous connectivity because it is available at the institute.

Turn it into a study

Measure connectivity in a rural facility across a week and design to what you find.

Fail-safe and fail-informative design

A device must fail into a safe state, and it must tell the user it has failed. Silent failure is the dangerous kind.

Where people go wrong

Designing for reliability without designing for failure.

Turn it into a study

Review failure modes of one device and whether each is detectable by the user.

Regulatory pathway for medical devices

Devices are classified by risk, and the evidence required rises with the class. The pathway must be planned before the design is frozen.

Where people go wrong

Building first and discovering the evidence requirement afterwards.

Turn it into a study

Classify a proposed device and map its regulatory requirements.

Usability testing with real users

Testing with actual users in the actual setting reveals failures that no engineering review finds.

Where people go wrong

Testing with colleagues who already understand the device.

Turn it into a study

Run a usability test with untrained users and count errors.

Measurement uncertainty

Every measurement has an uncertainty, and a device that reports more digits than its uncertainty supports is lying politely.

Where people go wrong

Displaying precision the instrument does not possess.

Turn it into a study

Compute measurement uncertainty for one device and compare with its display resolution.

Interoperability standards

Devices and records that cannot exchange data create silos, and standards exist precisely because every manufacturer would otherwise invent its own.

Where people go wrong

Building a system with a proprietary format and discovering the integration cost later.

Turn it into a study

Map the data standards used by systems in your hospital.

Lifecycle and maintenance

Most medical equipment in low-resource hospitals fails for want of maintenance, consumables or training rather than from design faults.

Where people go wrong

Measuring a donation programme by devices delivered.

Turn it into a study

what fraction of donated equipment in your hospital is still working.

Medical devices are regulated by risk class

Devices are classified by the harm they could cause, and the evidence, quality system and oversight required rise with the class, which must be known before design is frozen.

Where people go wrong

Designing first and discovering the class later.

Turn it into a study

Classify a proposed device under Indian device rules.

India regulates devices as drugs

India brought medical devices under its drugs legislation with a phased notification of device categories, which means device regulation sits within a pharmaceutical framework.

Where people go wrong

Assuming device regulation mirrors other countries.

Turn it into a study

Review which device categories are currently notified in India.

Software can be a medical device

Software that informs clinical decisions is itself regulated as a device in many jurisdictions, which brings app development under medical regulation.

Where people go wrong

Treating a clinical app as ordinary software.

Turn it into a study

Determine whether a clinical app you know would be regulated.

Intended use defines the device

A device's regulatory classification, testing and labelling all follow from its stated intended use, so a claim added to marketing can change the whole regulatory position.

Where people go wrong

Adding clinical claims to marketing material casually.

Turn it into a study

Write an intended use statement for a device and trace its consequences.

Risk management is a documented process

Device development requires systematic hazard identification, risk estimation, mitigation and verification, documented in a file that regulators examine.

Where people go wrong

Managing risk through design review discussions.

Turn it into a study

Build a hazard analysis for a simple device.

Single fault conditions must be safe

A device must remain safe when any single component fails, which drives redundancy, alarms and fail-safe default states.

Where people go wrong

Designing for correct operation only.

Turn it into a study

Identify single fault conditions in a device and their consequences.

Electrical safety has patient-specific standards

Leakage current limits for equipment connected to patients are far stricter than for ordinary appliances, and applied parts in contact with the heart stricter again.

Where people go wrong

Applying general electrical standards to medical equipment.

Turn it into a study

Review leakage current limits by applied part classification.

Biocompatibility is tested, not assumed

Materials contacting tissue require testing for cytotoxicity, sensitisation and irritation appropriate to the contact duration and type.

Where people go wrong

Selecting a material by mechanical properties alone.

Turn it into a study

Review biocompatibility testing requirements for a contact duration.

Sterilisation must be validated for the device

A device's sterilisation method must be validated for that device, because the method affects materials and the geometry affects whether sterilant reaches every surface.

Where people go wrong

Assuming a standard cycle sterilises any device.

Turn it into a study

Review sterilisation validation requirements for a device design.

Reprocessing single-use devices

Reusing devices labelled single-use is common in low-resource settings, and it raises documented infection and performance risks that reprocessing protocols only partly address.

Where people go wrong

Reusing a single-use device without a validated protocol.

Turn it into a study

Review single-use device reuse practice in your hospital.

Human factors engineering prevents use error

Designing controls, displays and alarms around how people actually behave under pressure prevents errors that training cannot, and regulators now require usability engineering.

Where people go wrong

Treating use error as operator fault.

Turn it into a study

Conduct a use-error analysis of a device in your hospital.

Alarm design is a discipline

Alarm priority, sound, distinctiveness and threshold determine whether alarms are acted on, and poorly designed alarms produce desensitisation and harm.

Where people go wrong

Adding alarms to improve safety.

Turn it into a study

Count distinct alarm sounds in your intensive care unit.

Default settings are decisions

Most users never change a default, so the default setting is the setting, which makes choosing defaults a clinical rather than engineering decision.

Where people go wrong

Setting defaults for engineering convenience.

Turn it into a study

Review default settings on devices in your hospital.

Interoperability requires standards

Devices and records exchange data only through shared standards, and proprietary formats produce silos that are expensive to bridge later.

Where people go wrong

Procuring systems without interoperability requirements.

Turn it into a study

Review the data standards used by systems in your hospital.

Health data standards exist

Defined standards for clinical data exchange and terminology allow systems to share meaning rather than only data, and India's health stack specifies which to use.

Where people go wrong

Designing a data format from scratch.

Turn it into a study

Review the standards specified in India's digital health architecture.

India built a digital health architecture

A national framework specifying health identifiers, registries and consent-based data exchange has been established, which shapes how any Indian health application must be built.

Where people go wrong

Designing a health application without reference to the national architecture.

Turn it into a study

Review the architecture's components and what they require of applications.

The Digital Personal Data Protection Act

India's data protection law sets obligations for collecting, processing and securing personal data, including health data, with consequences for health applications.

Where people go wrong

Building a health application without reference to data protection law.

Turn it into a study

Review the Act's requirements applicable to a health application.

De-identification is harder than it looks

Removing names does not anonymise a dataset, because combinations of age, location and dates re-identify individuals, which is why anonymisation is assessed rather than assumed.

Where people go wrong

Treating name removal as anonymisation.

Turn it into a study

Attempt re-identification on a de-identified dataset.

Clinical validation differs from technical performance

A device that measures accurately may not improve any clinical outcome, and the two require different studies, which is the gap most health technology falls into.

Where people go wrong

Presenting technical accuracy as clinical evidence.

Turn it into a study

Design both a technical and a clinical study for one device.

Diagnostic accuracy studies have a reporting standard

Studies of follow a reporting checklist, because the design features that results are predictable and must be reported.

Where people go wrong

Reporting accuracy without the design details.

Turn it into a study

Appraise a against the checklist.

The reference standard limits the measurement

A new test cannot be shown more accurate than an imperfect against which it is judged, which is a structural limit on diagnostic studies.

Where people go wrong

Treating disagreement with the reference as error in the new test.

Turn it into a study

Identify the 's limitations for one test.

Bland-Altman compares methods

Agreement between two measurement methods is assessed by the differences across the range rather than by correlation, which can be high while agreement is poor.

Where people go wrong

Using correlation to show two methods agree.

Turn it into a study

Plot a comparison for two devices.

Calibration matters as much as discrimination

A prediction model can rank patients correctly and still give badly wrong probabilities, and calibration is routinely unreported.

Where people go wrong

Evaluating a model by its area under the curve alone.

Turn it into a study

Plot a calibration curve for a model on local data.

Class imbalance distorts accuracy

When a condition is rare, a model that predicts nobody has it achieves high accuracy, which is why , precision and recall are reported instead.

Where people go wrong

Reporting accuracy for a rare outcome.

Turn it into a study

Compute several metrics for an imbalanced classification problem.

Leakage inflates model performance

Information from the outcome leaking into the features, or the same patient appearing in training and test sets, produces performance that collapses in practice.

Where people go wrong

Splitting data randomly without checking for patient overlap.

Turn it into a study

a model's data split for leakage.

External validation is the real test

A model must be tested on data from a different time, place or population, because internal validation measures how well it fits the data it was built on.

Where people go wrong

Reporting cross-validated performance as validation.

Turn it into a study

Test a model on data from a different source.

Dataset shift breaks deployed models

Populations, equipment and practice change over time, so a model's performance degrades after deployment unless it is monitored and updated.

Where people go wrong

Deploying a model without performance monitoring.

Turn it into a study

Design a monitoring plan for a deployed model.

Algorithmic bias is measurable

Model performance can differ systematically between groups, and measuring it by subgroup is the only way to detect it, which has been demonstrated in widely used health algorithms.

Where people go wrong

Reporting overall performance only.

Turn it into a study

Measure a model's performance by subgroup on local data.

A health algorithm encoded cost as need

A widely used American algorithm used healthcare spending as a proxy for health need, which systematically under-identified Black patients because less was spent on them.

Where people go wrong

Choosing a convenient proxy for an outcome.

Turn it into a study

Examine the proxy outcomes used in an algorithm you know.

Explainability has levels

Feature importance, example-based explanation and inherently interpretable models offer different kinds of explanation, and which is needed depends on who must act on it.

Where people go wrong

Treating explainability as a single property.

Turn it into a study

Compare explanation types for one clinical model.

Regulatory approval of adaptive algorithms

Algorithms that learn after deployment raise the question of what exactly was approved, which regulators address with predetermined change control plans.

Where people go wrong

Updating a deployed clinical algorithm without a change plan.

Turn it into a study

Review how regulators handle continuously learning algorithms.

Low-cost design is a discipline

Designing to a tenth of the cost requires removing functions, choosing different materials and rethinking the problem rather than cheapening components.

Where people go wrong

Reducing cost by substituting cheaper parts.

Turn it into a study

Redesign a device function at a tenth of its cost.

The ventilator built in a hurry

Rapidly designed emergency ventilators during the pandemic revealed that the hard parts were not the mechanism but the control, alarms, monitoring and validation.

Where people go wrong

Equating a working mechanism with a usable device.

Turn it into a study

Compare an emergency design with a certified ventilator's requirements.

Donated equipment often fails

A large share of donated medical equipment in low-income hospitals is not working, usually for want of consumables, spares, manuals or training rather than from breakage.

Where people go wrong

Measuring a donation by equipment delivered.

Turn it into a study

working status of donated equipment in your hospital.

Maintenance capacity is the limiting factor

Biomedical engineering capacity determines whether equipment keeps working, and hospitals often have more equipment than they can maintain.

Where people go wrong

Procuring equipment without maintenance capacity.

Turn it into a study

Count equipment items per biomedical technician in your hospital.

Total cost of ownership exceeds purchase price

Consumables, calibration, spares, service contracts and power over a device's life usually exceed its purchase price, which changes procurement comparisons.

Where people go wrong

Comparing devices by purchase price.

Turn it into a study

Model total cost of ownership for two device options.

Power quality damages equipment

Voltage fluctuation and interruption damage medical electronics, and protection and backup are part of the installation rather than optional extras.

Where people go wrong

Installing equipment without power conditioning.

Turn it into a study

Measure power quality where sensitive equipment is installed.

Design for the actual environment

Dust, humidity, temperature, voltage and untrained operators are design requirements in Indian field conditions, and laboratory validation does not test them.

Where people go wrong

Validating in a laboratory and deploying to a field setting.

Turn it into a study

Test a device against each environmental constraint.

Connectivity cannot be assumed

Intermittent or absent connectivity means devices must function and store data offline and synchronise later, which is an architectural decision rather than a feature.

Where people go wrong

Designing a cloud-dependent clinical application.

Turn it into a study

Measure connectivity at a rural facility across a week.

Battery life determines field usability

A device that cannot run a full clinic session on one charge is unusable in a field setting, which makes power budgeting a primary design constraint.

Where people go wrong

Specifying battery life from laboratory usage patterns.

Turn it into a study

Measure device power consumption under field usage.

Smartphones are medical instruments

Phone cameras, microphones, accelerometers and screens have been used for fundus imaging, auscultation, gait analysis and vision testing, which makes a widely owned device a diagnostic platform.

Where people go wrong

Treating a phone as a data entry device.

Turn it into a study

Identify a phone sensor suited to a clinical measurement.

Phone cameras vary enormously

An application relying on camera images must handle wide variation in sensors, processing and lighting between phone models, which is a validation problem rather than a nuisance.

Where people go wrong

Validating a camera-based application on one phone model.

Turn it into a study

Test an image-based application across several phone models.

Wearables measure continuously and imprecisely

Consumer wearables detect patterns over time that clinic measurements miss while being individually less accurate, which changes what questions they can answer.

Where people go wrong

Treating a wearable reading as a clinical measurement.

Turn it into a study

Compare wearable and clinical measurements over a week.

Screening with a wearable creates patients

A wearable that detects asymptomatic arrhythmia in a young population generates many false positives and incidental findings, which is a screening programme without a programme.

Where people go wrong

Treating consumer screening features as clinically neutral.

Turn it into a study

Review the atrial fibrillation screening studies using wearables.

Telemedicine needed guidelines

India issued telemedicine practice guidelines in 2020 defining when consultation, prescription and follow-up may occur remotely, which gave legal clarity to existing practice.

Where people go wrong

Practising telemedicine without reference to the guidelines.

Turn it into a study

Review the guidelines' requirements for prescribing remotely.

Asynchronous beats synchronous for some care

Store-and-forward consultation, where images and history are sent and reviewed later, works better than video for conditions where the information is visual and the timing is flexible.

Where people go wrong

Designing all telemedicine as video consultation.

Turn it into a study

Compare asynchronous and video consultation for one condition.

Implementation determines effect

Most digital health interventions fail in implementation rather than in concept, and implementation research examines workflow, incentives and training rather than the technology.

Where people go wrong

Evaluating a digital intervention by its technical function.

Turn it into a study

Design an implementation evaluation for a digital tool.

The pilot that never scaled

Indian digital health is full of successful pilots that never scaled, usually because scaling required integration, financing and workforce changes the pilot avoided.

Where people go wrong

Judging an intervention by pilot results.

Turn it into a study

Trace what happened to a digital health pilot in your state.

Open source in health technology

Open hardware and software designs for medical devices allow local manufacture and adaptation, and they shift the constraint from licence to manufacturing capability.

Where people go wrong

Assuming medical device designs must be proprietary.

Turn it into a study

Review an open source medical device project.

Molecules for MNR College of Engineering & Technology
The proteins behind these concepts are in the 3D Lab, where every one can be turned in 3D. Open the molecules for MNR College of Engineering & Technology →

Study ideas for MNR College of Engineering & Technology

Built from what this department works on. Take one to your guide before you collect anything.

Dry labThe performance of machine learning models for chest radiograph screening: a systematic review and meta-analysisDesign: Why it is worth doing here: A review needs only method, time and two reviewers.You would need: , two independent reviewers for screening and extraction, and a search built with a librarian.Start it in the Research guide
LaboratoryAgreement between a low-cost prototype device and manual counting by a clinician for measuring respiratory rate among patients attending the outpatient departmentDesign: Why it is worth doing here: A build-and-validate project with a clinical partner already on campus.You would need: , both methods applied to every participant in the same sitting, , and rather than correlation.Start it in the Research guide
SurveyImage quality in smartphone photographs of skin lesions among patients attending the skin outpatient department: a cross-sectional studyDesign: Why it is worth doing here: Joins the engineering side to a dermatology dataset that already exists.You would need: , a , a validated measurement tool, and written consent from every participant.Start it in the Research guide

Looking at another subject? Every subject is open to everyone, and a concept read through two departments is the fastest way to see how they join. Browse the other 37 subjects.